Linical Accuracy of a Continuous Glucose Monitoring System With an Advanced Algorithm

Introduction

Glycemic control remains suboptimal in pediatric patients with diabetes, with fewer than 25% of youth with type 1 diabetes (T1D) achieving glycemic targets of hemoglobin A1c (HbA1c) of <7.5%.1–4 Use of advanced diabetes technologies, like continuous glucose monitoring (CGM) devices, has been associated with improved glycemic control5–8 in some populations, like adults with T1D, but not universally, as pediatric samples have not experienced uniform benefits.5,9–14 Improved glycemic outcomes with CGM have been associated with consistent use of CGM.5,15,16 Thus, the lack of glycemic benefit of CGM among youth with T1D has been attributed to inconsistent use. In particular, only 30% of 15–24-year-old patients and 50% of 8–14-year-old patients in the JDRF CGM study wore CGM devices consistently for 6 or more days per week for the 6-month study duration compared with 86% of those 25 years of age and older.5 The latter group experienced a significant improvement in HbA1c compared with the control group using blood glucose (BG) monitoring alone. Notably, the devices used in these initial trials used early-generation technologies with accuracy, reliability, and usability issues.

The lack of consistent CGM use in the pediatric population has been attributed to several barriers associated with CGM. These barriers have been particularly evident in some of the first clinically available CGM devices, with barriers, for example, related to suboptimal accuracy and highly variable performance.17,18 These inaccuracies were attributable to many causes, including the time lag between interstitial and capillary glucose measurements and suboptimal algorithms that were used to derive the real-time glucose levels displayed on the CGM receiver. These challenges are especially evident in the pediatric population, who experience much greater glycemic variability and wider glycemic excursions than their adult counterparts with T1D.19,20 Indeed, it is not surprising that durability of CGM use has been modest in the pediatric population. The Type 1 Diabetes Exchange (T1DX) has documented limited CGM use, especially with the older CGM devices, as only 2–3% of pediatric patients 2–5, 6–12, and 13–17 years of age were using CGM in 2010–2012.2 A recent publication from the T1DX reports increased although still very limited uptake of CGM in pediatric patients, with only 13% of 2–5 year olds, 8% of 6–12 year olds, and 5% of 13–17 year olds using CGM.4

One of the main reasons for CGM discontinuation among the early adopting pediatric patients with T1D and their families related to sensor inaccuracies, leading to frustration when CGM and BG meter readings differed and when so-called "nuisance" CGM alerts and alarms were frequent.17,21 The temporal increase in CGM uptake documented in the T1DX coincides with improved accuracy of CGM devices,22 for example, related to the availability of the Dexcom G4® Platinum compared with the earlier Dexcom STS™ or Seven® Plus systems.23 Notably, CGM users compared with nonusers had better glycemic control in the T1DX, independent of their form of insulin delivery, and this pattern was evident in the baseline registry data from 2010–2012 as well as in the most recently reported annual update covering the period 2013–2014.2,4 Therefore, improved CGM algorithm performance leading to better accuracy of CGM devices could likely lead to greater uptake and persistent use of CGM in the pediatric population, potentially yielding greater glucose time in range and improved glycemic control.

The current study aims to compare the CGM performance of two recently available CGM systems, the Dexcom G4 Platinum RT-CGM (G4P) and Dexcom G4 Platinum with Software 505 algorithm (SW505), in one of the largest pediatric accuracy reports to date, including more than 250 youth 2–17 years of age with T1D. The SW505 algorithm is the same algorithm as the one used in the recently released Dexcom G5 Mobile™ CGM system for pediatric patients, 2 years of age and older, as well as adult patients. It is important to study critically the accuracy and performance as well as safety for regulatory approval, but it also necessary to report to the clinical community substantive improvements in CGM performance in order to advance efforts at increasing CGM uptake aimed at optimizing control in the very vulnerable pediatric population in need of improved glycemic outcomes in order to preserve health and prevent both short- and long-term complications of diabetes.

Research Design and Methods

This report includes two studies of CGM accuracy in pediatric patients with diabetes, 2–17 years of age. Both studies were short-term 1-week open-label single-arm multicenter studies including youth treated with multiple daily injections or insulin pump therapy. In the first study of the G4P device involving six centers (Study 1), youth could have T1D or type 2 diabetes (only one of 176 patients studied had type 2 diabetes). In the second study of the SW505 algorithm involving five centers (Study 2), all youth had T1D. Exclusion criteria included hematocrits beyond the range recommended by the study glucose meters, pregnancy, hypoglycemic unawareness (other than that usually expected for toddlers with diabetes), need for treatment with acetaminophen, and any significant illness that would pose a risk to the patient or to the staff handling the blood specimen.

In Study 1, participants wore two CGM systems simultaneously for a 7-day sensor wear period (amounting to 168 h), with one receiver providing real-time data and the other masked. In Study 2, participants wore a single unmasked sensor for the 7-day sensor wear period with CGM data displayed in real time. Sensors were inserted in the abdomen and/or upper buttocks by the patients or parents/guardians after self-training using a tutorial and/or one-on-one training by study staff.

Subjects in both studies were required to use a study-provided BG meter (the LifeScan [Milpitas, CA] OneTouch® Verio® IQ in Study 1 and the Bayer [Whippany, NJ] Contour™ Next USB BG meter in Study 2) and study-provided BG test strips for all BG measurements performed during sensor wear. In both studies, participants were asked to perform a minimum of seven fingersticks per day for home use for calibration (performed twice daily per labeling recommendations), comparative purposes, and all insulin dose selections as well as any other diabetes management decisions. All subjects avoided use of acetaminophen during the sensor wear period and for at least 24 h prior to sensor insertion. In both studies, participants were required not to inject insulin or wear an insulin pump insertion set within 3 inches of the sensor sites during wear. During home use, although the CGM data were displayed on the receiver screen, participants and families were to base all diabetes management decisions on results from the BG meter.

In both studies, subjects 6–17 years of age participated in one in-clinic session on either Day 1, 4, or 7 of sensor wear to allow for comparison of both the G4P and SW505 sensor glucose measurements with a reference glucose measurement (YSI BG analyzer; YSI, Yellow Springs, OH) obtained every 15 ± 5 min using arterialized venous blood and with glucose meter results using fingerstick capillary samples obtained every 30 ± 5 min. For youth 2–5 years of age, only fingerstick capillary samples were obtained every 30 ± 5 min. The CGM systems were calibrated using BG meter results at the start of the clinic session. During the in-clinic sessions, all receivers were masked, avoiding any display of sensor glucose results.

The preschool-aged subjects (2–5 years of age) participated in a 4-h clinic session in both studies. School-aged subjects (6–12 years of age) participated in the clinic session for up to 6 h. Teen subjects (13–17 years of age) participated in the clinic session for up to 6 h in Study 1 and for up to 12 h in Study 2. Those 6–17 years of age underwent intravenous catheterization of the dorsal hand, lower arm, or antecubital region for venous blood sampling for YSI plasma glucose determinations. The intravenous site was covered with a heating pad in order to arterialize the blood to more closely match the capillary BG concentrations. YSI sampling did not exceed 3 mL/kg of each subject's body weight. In the teen sample in Study 2, glucose levels during the in-clinic session were manipulated under close supervision according to protocol guidelines in efforts to achieve glucose levels across the range of sensor performance (40–400 mg/dL) by either adjusting the timing of meals (delaying meals after insulin administration to induce controlled hypoglycemia) or the timing of insulin administration (delaying insulin dosing with food intake to induce hyperglycemia). For youth of all ages in Study 1 and for those 2–12 years of age in Study 2, the participants checked their BG levels, took insulin, and ate as per their usual practice. At the end of the clinic session, study staff unmasked the receivers, and all receivers and glucose meters were downloaded using a sponsor-provided clinical laptop.

For both studies, on Day 7, participants returned to the clinic, removed the sensors themselves (or the sensors were removed by parents/guardians, as applicable), and returned the CGM systems and BG meters to the study staff. Study staff assessed adverse events related to study procedures, device use, and skin irritation. Staff carefully inspected the skin at the sites of sensor insertions and used the Draize scale to grade skin irritation.24 Sensors were inspected by the study staff and the sponsor to assess any sensor breakoff during use.

Both study protocols were approved by the site's Institutional Review Board or a central Institutional Review Board according to local policy, and all participants and parents/guardians provided signed assent (as appropriate for the youth's age) and informed consent, respectively, prior to performance of any study procedures. Both studies were reviewed and approved by the Food and Drug Administration prior to performance through the Investigational Device Exemption process.

Statistical methods and analyses

All data are presented as mean ± SD values or percentages unless otherwise indicated. Parametric (unpaired t tests) and nonparametric (Wilcoxon tests) were used as indicated by the sample distribution of the means. For comparisons of categorical variables, χ2 tests were used.

The real-time CGM values were compared with the temporally matched glucose values from the reference YSI and BG meter values. The mean absolute relative difference (MARD), as well as median absolute relative difference (ARD), in percentages and the proportion of the CGM system values that were within (±) 20% of the relative difference from the reference value at glucose levels >80 mg/dL and within (±) 20 mg/dL of absolute difference at glucose levels ≤80 mg/dL (hereafter referred to as %20/20) were used to evaluate the overall accuracy performance of the CGM devices. Similarly, performances of the G4P and SW505 systems were assessed within glucose ranges of 40–60 mg/dL, 61–80 mg/dL, 81–180 mg/dL, and >180 mg/dL as well as within each day of sensor use, from Day 1 through Day 7. The mean absolute difference (MAD), as well as median absolute difference, was also used to assess performance in the hypoglycemic range, 40–80 mg/dL.

Modified Bland–Altman plots25 were used to depict the data distribution and bias between the CGM and the reference glucose determinations. Error grid analyses, including the Clarke Error Grid26 and Parkes Error Grid,27 were used to quantify the clinical accuracy of the CGM devices. CGM diagnostic features were evaluated in the hypoglycemic range ≤80 mg/dL and in the hyperglycemic range ≥240 mg/dL by assessing concordance of CGM values within 15 min of the reference YSI results at these levels. Rate of both low and high false CGM alerts was also assessed based on CGM readings in the hypoglycemic or hyperglycemic ranges, respectively, when the matched YSI values within 15 min were not out of range. All analyses were performed using SAS software version 9.3 (SAS Institute, Inc., Cary, NC). P values of <0.05 were considered statistically significant.

Results

Participant characteristics and sensor usage

Study 1 consisted of 176 subjects, with 29 in the 2–5-year age group, 69 in the 6–12-year age group, and 78 in the 13–17-year age group, enrolled from six clinical centers in the United States. Study 2 included 79 subjects, with 16 in the 2–5-year age group, 17 in the 6–12-year age group, and 46 in the 13–17-year age group, enrolled from five clinical centers in the United States. Almost all patients had T1D, with an average duration about 5 years; the majority received insulin pump therapy. Mean HbA1c values were 8.2 ± 1.3% and 8.5 ± 1.5% in Studies 1 and 2, respectively. In Study 1, 40% of participants had previous exposure to CGM, whereas only 13% used CGM devices on a routine basis; in Study 2, 57% had previous exposure to CGM, whereas only 19% used it on a routine basis. Characteristics of the youth participants in Studies 1 and 2 are displayed in Table 1.

Table 1. Patient Characteristics in Studies of the Dexcom G4 Platinum Continuous Glucose Monitoring System and the G4 Platinum with the Software 505 Algorithm

Study sample
G4P (n = 176) SW505 (n =79)
Age (years) 11.4 ± 4.2 12.2 ± 4.6
Diabetes duration (years) 4.8 ± 3.7 5.6 ± 4.2
z-BMI
 Mean ± SD 0.5 ± 1.0 0.7 ± 0.7
 Range −4.7 to 2.6 −1.4 to 2.2
SMBG (times/day) 6.7 ± 2.9 6.7 ± 2.3
HbA1c (%) 8.2 ± 1.3 8.5 ± 1.5
Sex (% male) 57% 52%
Race (% white) 94% 96%
Type 1 diabetes (%) 99%a 100%b
Pump use (%) 72% 60%
Previous RT-CGM use (%) 40% 57%
Frequent (50% of time) RT-CGM use (%) 13% 19%
Age group (years) [n (%)]
 2–5 29 (16%) 16 (20%)
 6–12 69 (39%) 17 (22%)
 13–17 78 (44%) 46 (58%)

The expected number of sensors for use in both studies was 431 (176 × 2 sensors/subject in Study 1 and 79 × 1 sensor/subject in Study 2). Nearly all participants (244 out of 255, 96%) succeeded in inserting and wearing the originally placed sensors; only 11 subjects required sensor replacements. In total, 443 sensors were issued, indicating that 12 sensors were replaced among these 11 study subjects. Data from only six sensors out of 431 (1%) were excluded because of multiple reasons (e.g., deployment issues, inadvertent switch of transmitters between sib-pairs). Thus data were available from a total of 425 sensors in both studies. Of these, 367 sensors (86%) lasted through Day 7; 412 sensors (97%) lasted more than 4 days.

CGM performance

For assessment of CGM accuracy against YSI glucose measurements during the in-clinic sessions, there were 2,922 paired results (CGM and YSI temporally matched) in Study 1 and 2,262 paired results in Study 2 (Table 2). The overall MARD was 17% in Study 1 of the G4P, with a significant improvement in the MARD to 10% in Study 2 with the SW505 (P < 0.0001). Median ARD was 14% in Study 1 and 8% in Study 2. For assessment of CGM accuracy against self-monitoring of BG (SMBG) meter results during the 1 week of sensor wear, there were 16,318 paired results (CGM and meter temporally matched) in Study 1 and 4,262 paired results in Study 2. The overall MARD was 15% in Study 1 and 13% in Study 2 (P < .00001). The median ARD was 11% in Study 1 and 10% in Study 2. For both the G4P and SW505, CGM accuracy improved after Day 1 of sensor use (Fig. 1). For CGM accuracy against the YSI, MARD improved from 21% on Day 1 to 16% on Day 4 and then to 15% on Day 7 for the G4P; MARD improved from 13% on Day 1 to 8% on Day 4 and then to 10% on Day 7 for the SW505. For CGM accuracy against SMBG, MARD improved from 19% on Day 1 to 12% on Day 7 for the G4P; MARD improved from 15% on Day 1 to 11% on Day 7 for the SW505.

FIG. 1. 

FIG. 1. Mean absolute relative difference (MARD) (%) comparison of continuous glucose monitoring against (a) reference YSI across in-clinic study Days 1, 4, and 7 and (b) self-monitoring of blood glucose across home use Days 1–7. For continuous glucose monitoring accuracy against the YSI, MARD improved from 21% on Day 1 to 16% on Day 4 and then to 15% on Day 7 for the G4 Platinum (G4P); MARD improved from 13% on Day 1 to 8% on Day 4 and then to 10% on Day 7 for the Software 505 algorithm (SW505). For continuous glucose monitoring accuracy against self-monitoring of blood glucose, MARD improved from 19% on Day 1 to 12% on Day 7 for the G4P; MARD improved from 15% on Day 1 to 11% on Day 7 for the SW505.

Table 2. Continuous Glucose Monitoring (CGM) Performance Accuracy During Clinic (CGM vs. Reference YSI) and Home (CGM vs. Self-Monitoring of Blood Glucose) Use

CGM versus YSI CGM versus SMBG
G4P SW505 G4P SW505
Number of matched pairs 2,922 2,262 16,318 4,264
Mean/median ARD (%) 17/14 10/8 15/11 13/10
CEG Zone A (%)/A + B (%) 68/98 90/99 75/98 83/98
PEG Zone A (%)/A + B (%) 79/99 93/100 80/99 86/100
%20/20/%30/30 (%) 68/85 91/96 76/89 84/94
Within CGM ranges
 40 ≤ CGM ≤60 mg/dLa
  Number of matched pairs 19 86 487 240
  Mean/median AD (mg/dL) 19/9 16/13 24/18 17/14
 60 < CGM ≤80 mg/dLa
  Number of matched pairs 76 142 1,340 399
  Mean/median AD (mg/dL) 13/11 12/8 17/11 14/10
 80 < CGM ≤180 mg/dL
  Number of matched pairs 1,155 805 7,084 1,650
  Mean/median ARD (%) 17/13 11/8 15/11 12/9
 CGM >180 mg/dL
  Number of matched pairs 1,672 1,229 7,407 1,975
  Mean/median ARD (%) 18/14 9/7 14/10 11/8
 CGM >250 mg/dL
  Number of matched pairs 724 608 3,604 964
  Mean/median ARD (%) 18/15 10/7 14/10 11/8

Similarly, the Clarke Error Grid results indicated superior clinical accuracy with the SW505 algorithm compared with the G4P in the comparison of CGM versus YSI glucose values (Fig. 2a and b). In Study 1, 68% of values fell in Zone A, and 98% fell in Zones A and B for the G4P; in Study 2, 90% of values fell in Zone A, and 99% fell in Zones A and B for the SW505 algorithm. In Study 1, for the comparison of CGM with SMBG values, 75% of values fell in Zone A, and 98% fell in Zones A and B for the G4P (Fig. 2c); in Study 2, 86% fell in Zone A, and 98% fell in Zones A and B for the SW505 (Fig. 2d). Assessment of CGM accuracy using the Parkes Error Grid (Fig. 3) yielded a similarly improved performance with the SW505 compared with the G4P, with greater percentages of CGM falling within the clinically accurate Zone A and the combined Zone A plus the benign error Zone B for CGM versus YSI (Fig. 3a and b) and CGM versus SMBG (Fig. 3c and d). Notably, 100% of the CGM values fell within Zones A and B for CGM versus YSI and for CGM versus SMBG with the SW505.

FIG. 2. 

FIG. 2. Clarke Error Grids for continuous glucose monitoring (CGM) versus reference YSI (a and b) and self-monitoring of blood glucose (SMBG) (c and d) for the G4 Platinum (a and c) and G4P Platinum with Software 505 algorithm (b and d). The Clarke error grid results indicate superior performance with the Software 505 algorithm compared with the G4 Platinum in the comparison of CGM versus reference YSI glucose values and CGM versus SMBG. For the G4 Platinum, 68% of values fell in Zone A, and 98% fell in Zones A and B; for the Software 505 algorithm, 90% of values fell in Zone A, and 99% fell in Zones A and B. For the comparison of CGM with SMBG values, 75% of values fell in Zone A, and 98% fell in Zones A and B for the G4 Platinum; 86% fell in Zone A, and 98% fell in Zones A and B for the Software 505 algorithm.

FIG. 3. 

FIG. 3. Parkes Error Grids for continuous glucose monitoring (CGM) versus reference YSI (a and b) and self-monitoring of blood glucose (SMBG) (c and d) for the G4 Platinum (a and c) and G4P Platinum with Software 505 algorithm (b and d). Assessment of CGM accuracy using the Parkes Error Grid yielded improved performance with the Software 505 algorithm compared with the G4 Platinum, with greater percentages of CGM values falling within the clinically accurate Zone A and the combined Zone A plus the benign error Zone B for CGM versus reference YSI and CGM versus SMBG. Notably, 100% of the CGM values fell with Zones A and B for CGM versus YSI and for CGM versus SMBG with the Software 505 algorithm.

The %20/20 was 68% for the G4P and 91% for the SW505 algorithm in the comparison of CGM with YSI; the %20/20 was 76% for the G4P and 84% for the SW505 algorithm in the comparison of CGM with SMBG values. The (±) 30%/(±) 30 mg/dL values displayed the same pattern of improved performance accuracy for the SW505.

CGM performance was also assessed across various CGM glucose ranges where the SW505 performance was also superior to the G4P. Although the MARD was highest in the hypoglycemic range for glucose values 40–80 mg/dL, inclusive, the MAD offers an alternative assessment of accuracy in this range. MARD values were 18% for G4P and 17% for the SW505 (comparison of CGM with YSI) and 21% for the G4P and 19% for the SW505 algorithm (comparison of CGM with SMBG). For the glucose values 40–60 mg/dL, the MAD values were 19 mg/dL and 16 mg/dL for the G4P and SW505, respectively, in the comparison of CGM with YSI; the MAD values were 24 mg/dL and 17 mg/dL for the G4P and SW505, respectively, in the comparison of CGM with SMBG. For the glucose values 61–80 mg/dL, the MAD values were 13 mg/dL and 12 mg/dL for the G4P and SW505, respectively, in the comparison of CGM with YSI; the MAD values were 17 mg/dL and 14 mg/dL for the G4P and SW505, respectively, in the comparison of CGM with SMBG. Across other glucose ranges, the SW505 performed substantially better than the G4P with MARDs of 9–12% for the SW505 compared with 14–18% for the G4P (Table 2).

Bland–Altman density plots depict the bias of CGM to YSI and of CGM to SMBG in mg/dL for the G4P and the SW505 (shown in Fig. 4). The bias was centered around 0 with higher density for the SW505 compared with the G4P for both CGM versus YSI and CGM for SMBG. The majority of the bias for the SW505 fell within the modified International Organization for Standardization area of %20/20.

FIG. 4. 

FIG. 4. Bland–Altman density plots of continuous glucose monitoring (CGM) versus reference YSI glucose (a and b) and self-monitoring of blood glucose (SMBG) (c and d) measurements for the G4 Platinum (a and c) and the G4 Platinum with Software 505 algorithm (b and d). In Bland–Altman density plots, depicting the bias of CGM to YSI and of CGM to SMBG (in mg/dL) for the G4 Platinum and the Software 505 algorithm, the bias was centered around 0, with higher density for the Software 505 algorithm compared with the G4 Platinum for both CGM versus YSI and CGM for SMBG. The majority of the bias for the Software 505 algorithm fell within the modified International Organization for Standardization area of %20/20.

CGM detection of hypoglycemia and hyperglycemia

The SW505 performed superiorly to the G4P with respect to detection of hypo- and hyperglycemia. With a low glucose alert of 80 mg/dL, CGM detected true hypoglycemia according to YSI measurements ≤80 mg/dL 55% of the time within 15 min with the G4P and 91% of the time within 15 min with the SW505. In this hypoglycemic range, there was a false alert rate of 34% with the G4P and 14% with the SW505. With a high glucose alert of 240 mg/dL, CGM detected true hyperglycemia according to YSI measurements ≥240 mg/dL 96% of the time within 15 min with the G4P and 94% of the time within 15 min with the SW505. In this hyperglycemic range, there was a false alert rate of 33% with the G4P and 12% with the SW505.

Safety issues

There were no serious adverse events or device-related serious adverse events for either the G4P or the SW505 among the pediatric patients in either study. There was no sensor break-off or infection at the site of sensor insertion. There was mild skin irritation in some patients in the adhesive area, occurring at a low rate.

Conclusions

This report describes the performance of the G4P CGM system as well as the improved performance of the G4P with the SW505 algorithm in pediatric patients with diabetes. The initial study (Study 1), evaluating the safety and efficacy of the G4P, provided data for the Food and Drug Administration approval of the device for adjunctive use in pediatric patients, 2 years of age and older, in February 2014. The second study (Study 2) evaluated the safety and efficacy of the G4P with the updated SW505 algorithm for use in pediatric patients, using the same algorithm as that used in the adult G4P device and the currently approved Dexcom G5 Mobile CGM system. Indeed, Study 2 provided the safety and efficacy data for the SW505 algorithm for Food and Drug Administration approval for use in pediatric patients 2 years of age and older. The G5 Mobile, with the same advanced algorithm, also received the European CE mark as of September 2015 for use in pediatric patients as young as 2 years old, based on the overall improved performance. In pediatric patients, the SW505 has an MARD of 10% compared with reference YSI values, closely matching the favorable MARD of 9% in adults with diabetes.28 Similarly, clinical accuracy based on error grid analyses yielded 98–100% of values falling within Zones A and B in the pediatric patients compared with 99.5% for adults, and %20/20 was 91% for the younger patients and 93% for the adults for the SW505 in the comparisons of CGM with reference YSI glucose levels.28 Notably, the SW505 algorithm's performance has yielded opportunities for automated insulin delivery in many research settings.29,30

Although precision was not evaluated in Study 2 using the 505 algorithm, in Study 1 of the G4P, the overall paired absolute relative difference was 10%, and the overall coefficient of variation was 7% (data not shown). In addition, precision according to sensor site (abdomen and buttocks) was also satisfactory, varying from 6.3% (both sensors on abdomen) to 7.5% (one sensor on abdomen and one on buttocks), with the coefficient of variation midway at 7% for two sensors on the buttocks. Additionally, sensor performance for the G4P and the SW505 was fairly consistent across the pediatric age range, including the 2–5 year olds, the 6–12 year olds, and the 13–17 year olds. Thus, precision and accuracy of CGM in the pediatric population with diabetes now approach those in adults with diabetes. In addition, device safety has been documented.

Notably, almost all patients studied had T1D (except for a single insulin-treated patient with type 2 diabetes in Study 1). Patients with T1D, in general, and pediatric patients, in particular, display substantially more glycemic variability than do adults.19,31 Despite the inherent glycemic challenges experienced by these young patients, the performance of the SW505 system is sufficient to provide opportunities to optimize glycemic control in the pediatric population, given its enhanced performance characteristics. Indeed, pediatric patients will benefit from opportunities to increase glucose time in range while reducing risk of severe hypoglycemia.

Previous studies of CGM in the pediatric population have yielded inconsistent results regarding the ability of CGM to improve glycemic control, measured as HbA1c, in the pediatric population. The most common observation to date regarding CGM efficacy among pediatric patients relates to the lack of durability of CGM use. Although many pediatric patients may initiate CGM, numerous studies have documented the lack of sustained use among these young patients.9–14 Discontinuation has been related to the perceived burden or hassle factors of CGM overwhelming any potential benefits.17,18,21,32–34 These burdens arise from the device's size and appearance, pain associated with insertion and wearing, frequent alerts and alarms, and inherent inaccuracies in earlier systems. The alerts and alarms, often referred to as "nuisance" alarms, become particularly annoying when the SMBG value, obtained to validate the CGM reading, does not match the CGM glucose value. Thus, there has been a need for improved accuracy of devices that can reduce this perceived nuisance factor. The G4P with SW505 offers greater accuracy and alert performance, thereby providing a real opportunity to minimize this nuisance factor.

In addition, the improved CGM performance in the hypoglycemic range, including values between 40 and 80 mg/dL, may be able to provide reassurance that glycemic control can improve in the absence of an increased risk of severe hypoglycemia. Indeed, earlier pediatric studies have cited fear of hypoglycemia as a likely mediator of the previous failure of CGM to improve glycemic control in pediatric patients, especially in the younger patients, as parents focus on eliminating lower glucose levels in exchange for overall improvements in glycemia and target HbA1c attainment.13,14,35

A previous series of studies evaluated implementing CGM along with insulin pump therapy compared with multiple daily insulin injections (MDI) with SMBG in pediatric and adults patients with T1D.36–38 In these studies, known as STAR 3 (Sensor-Augmented Pump Therapy for A1C Reduction 3), using the Paradigm® REAL-time system from Medtronic (Northridge, CA), the pediatric patients, 7–18 years of age, who were randomized to sensor-augmented pump therapy compared with those randomized to MDI had significantly lower HbA1c levels, and a greater proportion of patients reaching age-specific glycemic targets had reduced glycemic variability, without increased hypoglycemia.37 Notably, the children, 7–12 years of age, used CGM more often than the adolescents, 13–18 years of age, in the 1-year randomized controlled trial, although both age groups benefited from the sensor-augmented pump therapy. During the 6-month extension phase of the study, the MDI group crossed over to sensor-augmented pump therapy and experienced significant improvements in HbA1c levels.38

The pediatric component of STAR 3 demonstrated opportunities for CGM implemented along with pump therapy to benefit the pediatric population. On the other hand, in the JDRF CGM study, which used multiple CGM devices, CGM use in the pediatric patients was not clearly associated with benefit in the younger patients, although adults derived significant benefit in that trial.5 Thus, in order for CGM to improve glycemic control without increased hypoglycemia in the pediatric population, CGM devices likely need to demonstrate improved performance or possibly be implemented with pump therapy. Additional opportunities may exist with the recent availability of wireless CGM data transmission to parents or other care providers, using systems such as the mySentry™ from Medtronic39 or Dexcom Share™.40

The studies described in the current report were short term and aimed at demonstrating safety and efficacy in order to achieve regulatory approval for use in the pediatric population. Longer-term studies are needed to assess whether the substantially improved performance of the SW505 algorithm results in greater uptake, sustained use, and improvements in glycemic control without an increase in severe hypoglycemia. Additionally, recent advancements in mobile communications with Bluetooth® (Bluetooth SIG, Kirkland, WA) connectivity that allows for transmission of CGM data directly to a smartphone, which eliminates the need to carry a separate receiver and then to transmit data directly to remote care providers, offer further opportunities to safely improve glycemic control in youth. These advances will likely promote increased uptake and durability of CGM use in pediatric patients; future studies are needed. Finally, use of the CGM data to make diabetes treatment decisions, including insulin dosing, will certainly reduce management burdens that are placed on patients and their families. Indeed, the Dexcom G4 Platinum with SW505 and the Dexcom G5 Mobile with SW505 have received the European CE mark for use in patients 2 years of age and older for nonadjunctive use, heralding in a potentially new era for diabetes management.

Acknowledgments

I would like to thank the participating sites, study staff, and the patients, acknowledging the site physicians and the biostatistician from Dexcom: Andrew Ahmann, MD, Portland, OR; Timothy Bailey, MD, Clinical Research, Advanced Metabolic Care and Research Institute, Escondido, CA; Bruce Buckingham, MD, Stanford University School of Medicine, Palo Alto, CA; Peter Chase, MD, Pediatrics, Barbara Davis Center for Diabetes, Denver, CO; Mark Daniels, MD, Children's Hospital of Orange County, Orange, CA; David Liljenquist, MD, Diabetes and Osteoporosis Center, Rocky Mountain, Idaho Falls, ID; Eva Tsalikian, MD, UIHC Department of Pediatrics, University of Iowa, Iowa City, IA; and Katherine Nakamura, PhD, Dexcom, San Diego, CA. The clinical studies were funded by Dexcom.

Author Disclosure Statement

L.L. reports consultancy for Johnson & Johnson, Eli Lilly, Sanofi, NovoNordisk, Roche Diagnostics, Dexcom, AstraZeneca, and Boehringer Ingelheim.

References

  • 1 Petitti DB, Klingensmith GJ, Andrews JS, Bell RA, Dabelea D, Imperatore G, Linder B, Loots B, Marcovina SM, Pihoker C, Rodriguez BL, Standiford DA, Waitzfelder B, Mayer-Davis EJ; SEARCH for Diabetes in Youth Study: Glycemic control in youth with diabetes: the SEARCH for Diabetes in Youth Study. J Pediatr 2009;155:668–672. Crossref, Medline, Google Scholar
  • 2 Beck RW, Tamborlane WV, Bergenstal RM, Miller KM, DuBose SN, Hall CA; T1D Exchange Clinic Network: The T1D Exchange clinic registry. J Clin Endocrinol Metab 2012;97:4383–4389. Crossref, Medline, Google Scholar
  • 3 Wood JR, Miller KM, Maahs DM, Beck RW, DiMeglio LA, Libman IM, Quinn M, Tamborlane WV, Woerner SE; T1D Exchange Clinic Network: Most youth with type 1 diabetes in the T1D Exchange Clinic Registry do not meet American Diabetes Association or International Society for Pediatric and Adolescent Diabetes clinical guidelines. Diabetes Care 2013;36:2035–2037. Crossref, Medline, Google Scholar
  • 4 Miller KM, Foster NC, Beck RW, Bergenstal RM, DuBose SN, DiMeglio LA, Maahs DM, Tamborlane WV; T1D Exchange Clinic Network: Current state of type 1 diabetes treatment in the U.S.: updated data from the T1D Exchange clinic registry. Diabetes Care 2015;38:971–978. Crossref, Medline, Google Scholar
  • 5 Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group: Continuous glucose monitoring and intensive treatment of type 1 diabetes. N Engl J Med 2008;359:1464–1476. Crossref, Medline, Google Scholar
  • 6 Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group: The effect of continuous glucose monitoring in well-controlled type 1 diabetes. Diabetes Care 2009;32:1378–1383. Crossref, Medline, Google Scholar
  • 7 Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group: Effectiveness of continuous glucose monitoring in a clinical care environment: evidence from the Juvenile Diabetes Research Foundation continuous glucose monitoring (JDRF-CGM) trial. Diabetes Care 2010;33:17–22. Crossref, Medline, Google Scholar
  • 8 Battelino T, Conget I, Olsen B, Schütz-Fuhrmann I, Hommel E, Hoogma R, Schierloh U, Sulli N, Bolinder J; SWITCH Study Group: The use and efficacy of continuous glucose monitoring in type 1 diabetes treated with insulin pump therapy: a randomised controlled trial. Diabetologia 2012;55:3155–3162. Crossref, Medline, Google Scholar
  • 9 Diabetes Research in Children Network (DirecNet) Study Group: Continuous glucose monitoring in children with type 1 diabetes. J Pediatr 2007;151:388–393. Crossref, Medline, Google Scholar
  • 10 Diabetes Research in Children Network (DirecNet) Study Group: FreeStyle Navigator continuous glucose monitoring system use in children with type 1 diabetes using glargine-based multiple daily dose regimens: results of a pilot trial. Diabetes Care 2008;31:525–527. Crossref, Medline, Google Scholar
  • 11 Diabetes Research in Children Network (DirecNet) Study Group: Prolonged use of continuous glucose monitors in children with type 1 diabetes on continuous subcutaneous insulin infusion or intensive multiple-daily injection therapy. Pediatr Diabetes 2009;10:91–96. Crossref, Medline, Google Scholar
  • 12 Chase HP, Beck RW, Xing D, Tamborlane WV, Coffey J, Fox LA, Ives B, Keady J, Kollman C, Laffel L, Ruedy KJ: Continuous glucose monitoring in youth with type 1 diabetes: 12-month follow-up of the Juvenile Diabetes Research Foundation continuous glucose monitoring randomized trial. Diabetes Technol Ther 2010;12:507–515. Link, Google Scholar
  • 13 Mauras N, Beck R, Xing D, Ruedy K, Buckingham B, Tansey M, White NH, Weinzimer SA, Tamborlane W, Kollman C; Diabetes Research in Children Network (DirecNet) Study Group: A randomized clinical trial to assess the efficacy and safety of real-time continuous glucose monitoring in the management of type 1 diabetes in young children (4 to <10 year olds). Diabetes Care 2012;35:204–210. Crossref, Medline, Google Scholar
  • 14 Tsalikian E, Fox L, Weinzimer S, Buckingham B, White NH, Beck R, Kollman C, Xing D, Ruedy K; Diabetes Research in Children Network (DirecNet) Study Group: Feasibility of prolonged continuous glucose monitoring in toddlers with type 1. Pediatr Diabetes 2012;13:294–300. Crossref, Medline, Google Scholar
  • 15 Deiss D, Bolinder J, Riveline JP, Battelino T, Bosi E, Tubiana-Rufi N, Kerr D, Phillip M: Improved glycemic control in poorly controlled patients with type 1 diabetes using real-time continuous glucose monitoring. Diabetes Care 2006;29:2730–2732. Crossref, Medline, Google Scholar
  • 16 Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group: Factors predictive of use and of benefit from continuous glucose monitoring in type 1 diabetes. Diabetes Care 2009;32:1947–1953. Crossref, Medline, Google Scholar
  • 17 Tansey M, Laffel L, Cheng J, Beck R, Coffey J, Huang E, Kollman C, Lawrence J, Lee J, Ruedy K, Tamborlane W, Wysocki T, Xing D; JDRF CGM Study Group: Satisfaction with continuous glucose monitoring in adults and youths with Type 1 diabetes. Diabet Med 2011;28:1118–1122. Crossref, Medline, Google Scholar
  • 18 Ramchandani N, Arya S, Ten S, Bhandari S: Real-life utilization of real-time continuous glucose monitoring: the complete picture. J Diabetes Sci Technol 2011;5:860–870. Crossref, Medline, Google Scholar
  • 19 Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group, Wilson DM, Xing D, Beck RW, Block J, Bode B, Fox LA, Hirsch I, Kollman C, Laffel L, Ruedy KJ, Steffes M, Tamborlane WV: Hemoglobin A1c and mean glucose in patients with type 1 diabetes: analysis of data from the Juvenile Diabetes Research Foundation continuous glucose monitoring randomized trial. Diabetes Care 2011;34:540–544. Crossref, Medline, Google Scholar
  • 20 Diabetes Control and Complications Trial Research Group: Effect of intensive diabetes treatment on the development and progression of long-term complications in adolescents with insulin-dependent diabetes mellitus: Diabetes Control and Complications Trial. J Pediatr 1994;125:177–188. Crossref, Medline, Google Scholar
  • 21 Polonsky WH, Hessler D: Perceived accuracy in continuous glucose monitoring: understanding the impact on patients. J Diabetes Sci Technol 2015;9:339–341. Crossref, Medline, Google Scholar
  • 22 Damiano ER, McKeon K, El-Khatib FH, Zheng H, Nathan DM, Russell SJ: A comparative effectiveness analysis of three continuous glucose monitors: the Navigator, G4 Platinum, and Enlite. J Diabetes Sci Technol 2014;8:699–708. Crossref, Medline, Google Scholar
  • 23 Christiansen M, Bailey T, Watkins E, Liljenquist D, Price D, Nakamura K, Boock R, Peyser T: A new-generation continuous glucose monitoring system: improved accuracy and reliability compared with a previous-generation system. Diabetes Technol Ther 2013;15:881–888. Link, Google Scholar
  • 24 Johnson AW, Goodwin BF: The Draize test and modifications. Curr Probl Dermatol 1985;14:31–38. Crossref, Medline, Google Scholar
  • 25 Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1:307–310. Crossref, Medline, Google Scholar
  • 26 Clarke WL, Cox D, Gonder-Frederick LA, Carter W, Pohl SL: Evaluating clinical accuracy of systems for self-monitoring of blood glucose. Diabetes Care 1987;10:622–628. Crossref, Medline, Google Scholar
  • 27 Parkes JL, Slatin SL, Pardo S, Ginsberg BH: A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose. Diabetes Care 2000;23:1143–1148. Crossref, Medline, Google Scholar
  • 28 Bailey TS, Chang A, Christiansen M: Clinical accuracy of a continuous glucose monitoring system with an advanced algorithm. J Diabetes Sci Technol 2015;9:209–214. Crossref, Medline, Google Scholar
  • 29 Russell SJ, El-Khatib FH, Sinha M, Magyar KL, McKeon K, Goergen LG, Balliro C, Hillard MA, Nathan DM, Damiano ER: Outpatient glycemic control with a bionic pancreas in type 1 diabetes. N Engl J Med 2014;371:313–325. Crossref, Medline, Google Scholar
  • 30 Kovatchev BP, Renard E, Cobelli C, Zisser HC, Keith-Hynes P, Anderson SM, Brown SA, Chernavvsky DR, Breton MD, Mize LB, Farret A, Place J, Bruttomesso D, Del Favero S, Boscari F, Galasso S, Avogaro A, Magni L, Di Palma F, Toffanin C, Messori M, Dassau E, Doyle FJ3rd: Safety of outpatient closed-loop control: first randomized crossover trials of a wearable artificial pancreas. Diabetes Care 2014;37:1789–1796. Crossref, Medline, Google Scholar
  • 31 Xing D, Kollman C, Beck RW, Tamborlane WV, Laffel L, Buckingham BA, Wilson DM, Weinzimer S, Fiallo-Scharer R, Ruedy KJ; Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group: Optimal sampling intervals to assess long-term glycemic control using continuous glucose monitoring. Diabetes Technol Ther 2011;13:351–358. Link, Google Scholar
  • 32 Diabetes Research in Children Network (DirecNet) Study Group: Youth and parent satisfaction with clinical use of the Glucowatch G2 Biographer in the management of pediatric type 1 diabetes. Diabetes Care 2005;28:1929–1935. Crossref, Medline, Google Scholar
  • 33 Diabetes Research in Children Network (DirecNet) Study Group: Psychological aspects of continuous glucose monitoring in pediatric type 1 diabetes. Pediatr Diabetes 2006;7:32–38. Crossref, Google Scholar
  • 34 Markowitz JT, Pratt K, Aggarwal J, Volkening LK, Laffel LM: Psychosocial correlates of continuous glucose monitoring use in youth and adults with type 1 diabetes and parents of youth. Diabetes Technol Ther 2012;14:523–526. Link, Google Scholar
  • 35 Tansey M, Weinzimer S, Beck R, Ruedy K, Cheng P, Tamborlane W, Kollman C, Mauras N, Fox L, Coffey J, White NH: Extended 6-month follow-up of a randomized clinical trial to assess the efficacy and safety of real-time continuous glucose monitoring in the management of type 1 diabetes in young children aged 4 to <10 years. Diabetes Care 2013;36:e63. Crossref, Medline, Google Scholar
  • 36 Bergenstal RM, Tamborlane WV, Ahmann A, Buse JB, Dailey G, Davis SN, Joyce C, Peoples T, Perkins BA, Welsh JB, Willi SM, Wood MA; STAR 3 Study Group: Effectiveness of sensor-augmented insulin-pump therapy in type 1 diabetes. N Engl J Med 2010;363:311–320. Crossref, Medline, Google Scholar
  • 37 Slover RH, Welsh JB, Criego A, Weinzimer SA, Willi SM, Wood MA, Tamborlane WV: Effectiveness of sensor-augmented pump therapy in children and adolescents with type 1 diabetes in the STAR 3 study. Pediatr Diabetes 2012;13:6–11. Crossref, Medline, Google Scholar
  • 38 Bergenstal RM, Tamborlane WV, Ahmann A, Buse JB, Dailey G, Davis SN, Joyce C, Perkins BA, Welsh JB, Willi SM, Wood MA; STAR 3 Study Group: Sensor-augmented pump therapy for A1C reduction (STAR 3) study: results from the 6-month continuation phase. Diabetes Care 2011;34:2403–2405. Crossref, Medline, Google Scholar
  • 39 Kaiserman K, Buckingham BA, Prakasam G, Gunville F, Slover RH, Wang Y, Nguyen X, Welsh JB: Acceptability and utility of the mySentry remote glucose monitoring system. J Diabetes Sci Technol 2013;7:356–361. Crossref, Medline, Google Scholar
  • 40 Dexcom. www.dexcom.com/dexcom-g4-platinum-share (accessed December 8, 2015 ). Google Scholar